DNA/DNR-AI-217 Simultaneous Sampling Differential Analog Input Board
Chapter 1 14
Introduction
Tel: 508-921-4600 www.ueidaq.com Vers: 4.5
Date: April 2013 DNx-AI-217 Chap1x.fm
© Copyright 2013
United Electronic Industries, Inc.
1.9.1 Analog Input
Ground
Connections
To avoid errors caused by common mode voltages on analog inputs, follow the
recommended grounding guidelines in Figure 1-6 below. Because all analog
input channels in AI-201/202/207/208/217/225 layers are isolated as a group,
you can connect layer AGND to the ground of the signal source and eliminate
the resistors shown in Figure 1-6 for floating differential input signals.
Figure 1-6. Recommended Ground Connections for Analog Inputs
1.10 CJC Data Raw CJC Voltage from the AI-217 may be represented as:
For example, if the voltage read from Channel 33 (the CJC channel) is 0.87, the
CJC Temperature is:
Input
Configuration
Type of Input
Floating Grounded
Typical Signal Sources:
Thermocouples
DC Voltage Sources
Instruments or sensors
with isolated outputs
Typical Signal Sources:
Instruments or sensors
with non-isolated outputs
Differential
Two resistors (10k <R< 100k) provide
return paths to ground for bias currents.
Single-Ended,
Ground
Referenced
+
_
Ret
AGnd
,
AInX
V
in
Sig
DNA-STP-AI-U
+
_
Ret
AGnd
DNA-STP-AI-U
AInX
V
in
Sig
+
_
Ret
AGnd
DNA-STP-AI-U
AInX
V
in
Sig
+
_
Ret
AGnd
DNA-STP-AI-U
AInX
V
in
Sig
grounds are at the same potential.
Add this connection to ensure that both
NOT RECOMMENDED
Temp. Scale Calculation CJC Temperature
Kelvin 0.87/0.00295 294.9 °K
Celsius 294.9 – 273.15 21.75 °C
Fahrenheit 1.8 x 21.75 + 32 71.75 °F
T
Kelvin
CJCVoltage
0.00295
--------------------------------=